ORIGINAL ARTICLE
Year : 2014  |  Volume : 17  |  Issue : 2  |  Page : 179-181
 

The sensitivity comparison of bar electrode and disposable ring electrode for recording of lateral femoral cutaneous nerve


1 Department of Medicine, Neuroscience Section, Virginia Tech Carilion School of Medicine, Roanoke, USA
2 Department of Neurology, George Washington University, Washington, DC, USA

Date of Submission16-Aug-2013
Date of Decision20-Sep-2013
Date of Acceptance24-Oct-2013
Date of Web Publication17-May-2014

Correspondence Address:
Ahmet Z. Burakgazi
Carilion Clinic, 3 Riverside Circle, Roanoke, Virginia - 24016
USA
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0972-2327.132621

Rights and Permissions

 

   Abstract 

Objectives: To compare the sensitivity of bar electrode and disposable ring electrode for recording of lateral femoral cutaneous (LFCN) Materials and Methods: A total of 23 subjects (13 females, 10 males, mean age: 49.6 ± 9.6 (range: 29-63) were recruited in the study. A total of 36 recordings were obtained with each electrode (with bar and disposable ring electrodes) from the subjects. The comparison of data was performed with percentages and student T-table test. Results: The response rate was 98% (35 out of 36 recordings) with bar electrode and 88% (32 out of 36 recordings) with disposable ring electrode. Although the sensitivity rate of bar electrode is slightly higher than of disposable ring electrode, there were no statistically significant differences in detecting the onset latency, peak latency, and amplitude of LFCN. Conclusion: The recording sensitivity of LFCN is higher with bar electrode than disposable ring electrode. However, disposable ring electrode can be used alternatively.


Keywords: Bar electrode, disposable ring electrode, lateral femoral cutaneous nerve, LCFN, meralgia paresthetica, nerve conduction studies


How to cite this article:
Burakgazi AZ, Russo M, Richardson PK. The sensitivity comparison of bar electrode and disposable ring electrode for recording of lateral femoral cutaneous nerve. Ann Indian Acad Neurol 2014;17:179-81

How to cite this URL:
Burakgazi AZ, Russo M, Richardson PK. The sensitivity comparison of bar electrode and disposable ring electrode for recording of lateral femoral cutaneous nerve. Ann Indian Acad Neurol [serial online] 2014 [cited 2019 Dec 13];17:179-81. Available from: http://www.annalsofian.org/text.asp?2014/17/2/179/132621



   Introduction Top


Meralgia Paresthetica (MG) is a peripheral nerve disorder caused by compression and/or dysfunction of lateral femoral cutaneous nerve (LFCN). [1],[2] LFCN is derived from the second and third lumbar roots and pure sensory nerve. [1],[2],[3] Thus, MG clinically manifests with paresthesias and numbness over the anterolateral part of the thigh. [2],[4]

The diagnosis of MG is usually based on clinical findings. However, lumbar radiculopathy rarely mimics MG. [5] Thus, electrodiagnostic testing such as nerve conduction studies (NCS) [1],[2],[3],[6] and somatosensory evoked potential [7],[8] is important to verify the diagnosis. Sensory NCS of LFCN is useful diagnostic tool. However due to the anatomic course of the LFCN, overlapping of LFCN response by shock artifact or motor response from the quadriceps, and obesity, the recording of LFCN with NCS is usually challenging for physicians and technicians. [4] Different techniques have been developed for obtaining the sensory nerve action potential (SNAP) of LFCN to improve sensitivity of NCS. [2],[9] The purpose of our study was to describe a method that provides larger SNAP amplitude of LFCN and the better rate of recorded SNAP with disposable ring electrode than bar electrode.


   Materials and Methods Top


A total of 23 subjects (13 females, 10 males, mean age 49.6 ± 9.6 (range: 29-63) were included in the study. The samples consisted of 23 normal subjects, without signs or symptoms of peripheral nerve disorders. A total of 36 recordings were obtained with each electrode (with bar and disposable ring electrodes) from the subjects. Bilateral recordings were performed in 13 subjects and unilateral recording was performed in 10 subjects.

NCS of the LFCN was performed by the same technician (initials: MR) with Cadwell Sierra electromyography machine using bar electrode (Shielded bar electrode with 5-pole, Cadwell, USA) and disposable ring electrodes [Wrap-around, silver/silver chloride ring electrodes; dimensions: 10 cm (4 in) long, 0.6 cm (0.25 in) wide, Cadwell, USA] for recording and a bipolar surface electrode for stimulation. Filter settings included frequency 30-20,000 Hz, sensitivity 5-20 μV per division, and sweep speed 2 ms/division. Stimulation intensity ranged from 6 mA to 42 mA with 0.1-ms duration. No averaging was performed. Temperature of the limb was maintained at 33°C. Electrophysiological testing of the LFCN was performed on each subject with two different electrodes with the same technique. Before performing the NCS test, the skin was cleaned with alcohol to decrease impedance. The stimulation electrode was placed 4 cm distal to the anterior superior iliac spine (ASIS). The pick electrodes [bar electrode [Figure 1]a] and disposable ring electrode [Figure 1]b were placed 10 cm distal to the stimulation site on a line connecting the ASIS to the lateral border of the patella. The subjects were asked to notify the operator when they felt stimulation to the anterolateral aspect of the thigh. [2] No averaging was used. Each SNAP was obtained at least three times in each trial to confirm the variability of the nerve conduction results between the different trials. [10] It took one to three trials to find the stimulation site. [10]
Figure 1: Lateral femoral cutaneous sensory stimulation either bar or disposable ring electrodes (a) Lateral femoral cutaneous with bar (b) Lateral femoral cutaneous with disposable ring electrodes

Click here to view


The comparison of data was performed with percentages and student T-table test. A P-value less than or equal to 0.05 was considered as a statistically significant.


   Results Top


The sensory NCS of LFCN results with the bar and disposable ring electrodes are summarized in [Table 1] and [Table 2]. Mean SNAP latency was measured at 1.59 ms [standard deviation (SD = 0.29)] with mean amplitude 7.76 μV (SD = 4.97) with bar electrodes, and at 1.62 ms (SD = 0.32) with mean amplitude 7.03 μV (SD = 4.23) with disposable ring electrodes. The response rate was 97% (35 out of 36recordings) with bar electrode and 88% (32 out of 36 recordings) with disposable ring electrode. In three recordings, LFCN response was unobtainable with disposable ring electrode but obtainable with bar electrode. In only one recording, LFCN response was unobtainable with bar electrode but obtainable with disposable ring electrode.
Table 1: The comparison of bar electrode and disposable ring electrode

Click here to view
Table 2: Recordings data from bar and ring electrodes

Click here to view



   Discussion Top


The use of NCS in the diagnosis of MG was first performed by Butler et al., [9] in 1974. In the first technique, surface electrodes are placed antidromically to record the action potential of the LFCN following stimulation at a site located 1 cm medial to the ASIS. Then, Russo et al., [2] described an alternative technique involving the placement of bar electrode antidromically to record the action potential of the LFCN following stimulation 4 cm distal to the ASIS. The response rate was 35% with the traditional technique, whereas it was 90% with the new alternative technique. The new technique is more sensitive then the traditional technique. [1] In el-Tantawi study, [7] the traditional technique was performed that may explain their low sensitivity rate (28.1%) with sensory NCS. We performed the new technique in our study. The sensitivity rate was 98% with bar electrode and 88% with disposable ring electrode. Overall, bar electrode was more sensitive than disposable ring electrode in recording LFCN.

The values of mean SNAP amplitude and latency have shown variation in different studies. In Laroy et al.,'s [10] study involving 58 healthy subjects, mean SNAP latency of LFCN was measured at 1.7 ms (SD = 0.23) with mean amplitude of 10.5 μV (SD = 0.4) by using surface disc electrodes for recording. [10] In their study, antidromical stimulation was performed with the bipolar electrode placed at 4.0 cm on vertical line distal to the ASIS. [10] In Lagueny et al.,'s [11] study mean value of SNAP amplitude was 12.10 μV with the antidromic study (A stimulating cathode was located 1.5 cm distal and medial to the ASIS with the anode 3 cm proximal to the cathode and recording electrodes were located on a straight line between the ASIS and the lateral border of the patella, with an interelectrode distance of 3 cm, the proximal being 12 cm distal to the ASIS) and 10.59 μV with orthodromic study (electrodes were left in place, but their use inversed and the study was performed by near-the-nerve or subcutaneous needle electrodes). [11] In Seror's study, [8] the mean SNAP amplitude was 16 ± 7 μV with orthodromic recording (stimulation was performed with a bipolar fixed distance 23 mm surface electrode, 12 cm distal to the subcutaneous needle recording electrodes (interelectrode distance, 25 mm). In our study, although the sensitivity rate of the bar electrode recording was higher than of the disposable ring electrode recording, there were no statistically significant differences in mean values of onset latency, peak latency, and amplitude of LFCN between bar and disposable ring electrodes.

In conclusion, the bar electrode recordings is more sensitive than disposable ones but is less reliable and less sensitive than antidromic recording performed by Lagueny and Laroy and than orthodromic recording performed with near-the-nerve or subcutaneous needle electrodes.

 
   References Top

1.Patijn J, Mekhail N, Hayek S, Lataster A, van Kleef M, Van Zundert J. Meralgia Paresthetica. Pain Pract 2011;11:302-8.  Back to cited text no. 1
    
2.Russo MJ, Firestone LB, Mandler RN, Kelly JJ Jr. Nerve conduction studies of the lateral femoral cutaneous nerve. Implications in the diagnosis of meralgia paresthetica. Am J Electroneurodiagnostic Technol 2005;45:180-5.  Back to cited text no. 2
    
3.Grossman MG, Ducey SA, Nadler SS, Levy AS. Meralgia paresthetica: Diagnosis and treatment. J Am Acad Orthop Surg 2001;9:336-44.  Back to cited text no. 3
    
4.Wiezer MJ, Franssen H, Rinkel GJ, Wokke JH. Meralgia Paraesthetica: Differential diagnosis and follow-up. Muscle Nerve 1996;19:522-4.  Back to cited text no. 4
    
5.Yang SN, Kim DH. L1 radiculopathy mimicking meralgia paresthetica: A case report. Muscle Nerve 2010;41:566-8.  Back to cited text no. 5
    
6.Sarala PK, Nishihara T, Oh SJ. Meralgia paresthetica: Electrophysiologic study. Arch Phys Med Rehabil 1979;60:30-1.  Back to cited text no. 6
[PUBMED]    
7.El-Tantawi GA. Reliability of sensory nerve-conduction and somatosensory evoked potentials for diagnosis of meralgia paraesthetica. Clin Neurophysiol 2009;120:1346-51.  Back to cited text no. 7
[PUBMED]    
8.Seror P. Lateral femoral cutaneous nerve conduction v somatosensory evoked potentials for electrodiagnosis of meralgia paresthetica. Am J Phys Med Rehabil 1999;78:313-6.  Back to cited text no. 8
[PUBMED]    
9.Butler ET, Johnson EW, Kaye ZA. Normal conduction velocity in the lateral femoral cutaneous nerve. Arch Phys Med Rehabil 1974;55:31-2.  Back to cited text no. 9
[PUBMED]    
10.Laroy V, Knoops P, Semoulin P. The lateral femoral cutaneous nerve: Nerve conduction technique. J Clin Neurophysiol 1999;16:161-3.  Back to cited text no. 10
    
11.Lagueny A, Deliac MM, Deliac P, Durandeau A. Diagnostic and prognostic value of electrophysiologic tests in meralgia paresthetica. Muscle Nerve 1991;14:51-6.  Back to cited text no. 11
    


    Figures

  [Figure 1]
 
 
    Tables

  [Table 1], [Table 2]



 

Top
Print this article  Email this article

    

 
   Search
 
  
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Article in PDF (356 KB)
    Citation Manager
    Access Statistics
    Reader Comments
    Email Alert *
    Add to My List *
* Registration required (free)  


    Abstract
   Introduction
    Materials and Me...
   Results
   Discussion
    References
    Article Figures
    Article Tables

 Article Access Statistics
    Viewed2497    
    Printed38    
    Emailed0    
    PDF Downloaded55    
    Comments [Add]    

Recommend this journal